Automatic freeze drying endpoint determination for vials application

Automatic freeze drying endpoint determination for vials application on Lyovapor™ L-300 Pro by pressure difference test, pressure rise test and temperature difference test

Automatic freeze drying endpoint determination for vials application

Freeze-drying is a self-contained process in which samples cannot be easily analysed in real-time to detect their residual moisture content. Process optimization, in particular the required process time to obtain a dry and stable product, often relies on a trials-and-error approach. In this application note, a combination of different process analytical technologies was applied to determine the endpoint of primary and secondary drying of mannitol solutions in a laboratory freeze-dryer (Lyovapor™ L-300). A setup with serum tubing vials on temperature-controlled shelves was used. The indirect tracking of drying progress via in-situ measurements of sample parameters enables an instant adjustment of process time in a running freeze-drying cycle. It contributes to faster parameter optimization according to the required residual moisture content of the product. Furthermore, these analytical techniques provide essential tools to monitor the process reproducibility.

Registrati

Paese
In futuro vorremmo inviarti altre informazioni, come ad esempio note applicative, guide, webinar e inviti a seminari, che potrebbero essere di tuo interesse. Per questo motivo ti chiediamo il consenso al contatto. Puoi trovare le informazoni sull'elaborazione e trattamento dei tuoi dati personali nella nostra politica sulla privacy