Vacuum Modules EasyVac/LabVac
![]() | The modules dedicated for specific functions: The EasyVac for automatic distillation, the LabVac for complete laboratory vacuum systems. |
EasyVac Limitations
What are the limitations of EasyVac regarding flask size and solvents?
The EasyVac module works fine with a lab-scale rotary evaporator and flask sizes between 100 ml and 4 litres. All typical solvents are precisely detected. If mixtures of solvents are used, it may happen that the second solvent is not detected and has to be distilled in a second step.
Manual adjustment EasyVac
During automatic distillation with the EasyVac, I want to speed up the process. How can I regulate the parameters?
By simply pressing the up / down button, the operator can adjust the pressure and thus the distillation process. The pressure remains on the adjusted level until the process is terminated.
EasyVac and display
Can I have the Vacuum Module V-801 EasyVac with display?
No, the EasyVac has no display. If the pressure display and the EasyVac function should be combined, we recommend the V-855 Vacuum Controller covering both features.
Drying and EasyVac
Can I use the EasyVac function for drying processes?
Typically, only a few amount of solvent is evaporated in a drying process. Thus, the EasyVac is not recommended.
LabVac compatibility
I have a Vacuum Pump V-7100 with B-719 controller. Can I use the LabVac module instead of the B-719?
No. The two types of controllers have different communication and protocols and are therefore not compatible.
LabVac and V-700 compatibility
Can I use the LabVac module also with the Vacuum Pump V-700?
Basically yes, However, as the typical application of the Vacuum Module V-802 LabVac is the combination of different vacuum users to one system, the Vacuum Pump V-700 is to small. The ideal combination is the Vacuum Pump V-710 and the LabVac module.